PDS_VERSION_ID = PDS3 ^PDF_DOCUMENT = "PFS_LWC_CALIB.PDF" PRODUCT_ID = PFS_LWC_CALIB RELEASE_ID = 0001 REVISION_ID = 0000 RECORD_TYPE = FIXED_LENGTH DATA_SET_ID = "MEX-M-PFS-2-EDR-EXT3-V1.0" PRODUCT_CREATION_TIME = 2010-12-07T12:01:20.887 OBJECT = PDF_DOCUMENT DOCUMENT_NAME = "CALIBRATION OF THE PLANETARY FOURIER SPECTROMETER LONG WAVELENGTH CHANNEL" PUBLICATION_DATE = 2010-12-07 DOCUMENT_TOPIC_TYPE = "CALIBRATION REPORT" INTERCHANGE_FORMAT = BINARY DOCUMENT_FORMAT = "ADOBE PDF" DESCRIPTION = "The Planetary Fourier Spectrometer (PFS) experiment on board the Mars Express mission has two channels covering the 1.2-5 microns ( SWC) and the 5-50 microns ( LWC ) . The Long Wawelength Channel (LWC) measures the thermal emission spectrum of Mars between 200cm-1 and 2000cm-1 with a spectral resolution of 1.4 cm-1, in absence of apodization. We present here the calibration of this channel and its performance . The instrument calibration has been performed on ground , before launch , in space during Near Earth Verification (NEV) measurements, and at Mars . Special attention has been given to the problem of microvibrations on board the spacecraft. In order to obtain correct results , the source-instrument-detector interaction is studied very accurately. The instrument variations during a pericentre pass impose a complex procedure for the LW channel calibration, but fortunately the procedure adopted seems to work well. Samples of the calibrated data are given (as single spectrum and as an average over a few spectra) to show the performance of the experiment and its scientific potentialities." END_OBJECT = PDF_DOCUMENT END