Version 2.0 Corrections and Updates =================================== The ROSINA team found errors in the V1.0 RTOF, DFMS, and COPS datasets resulting from various problems in the pipeline that produced the data. The pipeline was corrected and V2.0 of the ENG, CR5, EAR3, CR4B, and AST1 datasets were created using the improved pipeline. The specific changes are detailed below. The RTOF products were recreated to correct an error in extracting the science data from the raw telemetry. In the DFMS data, corrections were made to the HK parameters. The values of the following keywords have been recalculated and updated: ROSINA_DFMS_MEP_MOT_COUNTER ROSINA_DFMS_MEP_CEM_GND ROSINA_DFMS_MEPA_TEMPERATURE ROSINA_DFMS_MEPN_TEMPERATURE ROSINA_DFMS_MEPG_TEMPERATURE ROSINA_DFMS_MEP_H_TEMPERATURE In addition, an error in selecting and applying calibration curves was corrected and the data reprocessed. In the RTOF data, corrections were made to the HK parameters. The values of the following keywords have been recalculated and updated: ROSINA_RTOF_SCI_MC_OMC_SW_OPEN ROSINA_RTOF_SCI_MC_OMC_SW_OPEN ROSINA_RTOF_SCI_MC_OMC_SW_OPEN In addition, an error in selecting and applying calibration curves was corrected and the data reprocessed. In some files, the START_TIME and STOP_TIME values used were inadvertently taken from laboratory process files rather than spacecraft files. These times have been corrected. Note that in those cases where START_TIME is incorporated into the file name, this has resulted in file name changes from the previous version. Some COPS files have corrections in the HK table size. COPS has two gauges (Ram and Nude). The HK table size in the COPS/BG data has been corrected to be consistent with the status of the gauges. Specifically, when one gauge is not operating, its 3 offset keywords parameters are not present and the table size goes down from 72 to 69 bytes per row.