ROSETTA CIVA-P DATA CALIBRATION DESCRIPTION =========================================================================== Revisions --------- 2015-06-26 first issue, Purpose ------- This document describes the calibration of the CIVA-P data delivered in the CIVA level 2 data set. Introduction ============ The calibration of CIVA-P raw images is realized in the following steps (applied in that order): 1) Non-linearity correction The TH7888A detector presents a strong deviation from linearity when the signal is higher than ~ 80% of the saturation level. This saturation level depends on the camera, on the gain setting and on the temperature of the detector. A linearity correction transformation which includes the subtraction of the offest has therefore been applied to level 1 CIVA images which reach levels close to saturation (CIVA 1, CIVA 2, CIVA 3 and CIVA4). The linearity files providing the actual DN level from the measured DN level (1024 floating point values) are civa1scale.txt, civa2scale.txt, civa3scale.txt, civa4scale.txt. The offset was evaluated as 18.4 DN, 19 DN, 13 DN and 20.9 DN for these four cameras. Cameras CIVA 5, CIVA 6 and CIVA 7 have low DN values, hence the corrected values correspond to the subtraction of an offset of 17.8 DN, 20.5 DN and 14.5 DN for CIVA 5, CIVA 6 and CIVA 7 respectively. Regions of images from the CIVA 1, CIVA 2, CIVA 3 and CIVA 4 cameras are saturated. This situation is indicated in the image file by a value of the recovered DN equal to the maximum floating value of the transformation table "provided in ""civaNscale.txt"" (N=1 to 4). " Level 3 will contain following steps: Geometrical correction Photometric correction (Flat field)