Community Release Note for PACS Calibration Set v69
PACS calibration file set v69 incorporates the following differences from the previous community release, v65:
Spectrometer
-
New version of the BeamsPerSpaxel calibration files. These files contain the beams which were published on the
PACS Calibration WIKI in the file PCalSpectrometer_Beam_v5.tar.gz.
-
Provide fitted version of the Spectrometer BeamSize product.
The calfile spectrometer.beamSize version 1 only contained a few measures. The resulting curve is not really
smooth, and doesn't cover the whole wavelength range. A fitted version of those data is now provided,
interpolated over all wavelengths between 50 & 220 um. A polynomial fit of order 2 does a good job.
Photometer
Affected observations & processing
- Spectrometer: level-2 drizzle products and results of the specExtendedToPointCorrection task.
Updated calibration products
The following calibration products have been updated:
- PCalSpectrometer_BeamSize_FM_v3.fits
- PCalSpectrometer_BeamsPerSpaxelB2A_FM_v3.fits
- PCalSpectrometer_BeamsPerSpaxelB2B_FM_v3.fits
- PCalSpectrometer_BeamsPerSpaxelB3A_FM_v3.fits
- PCalSpectrometer_BeamsPerSpaxelR1_FM_v3.fits